Electrical Fast Transient Generator Calibration
We provide calibration services for Electrical Fast Transient Generators. This calibration provides 50 Ohm and 1,000 Ohm impedance network waveform data that is compared to the requirements listed in IEC 61000-4-4, basic standard.
This table describes our scope of accreditation for electrical fast transient generator calibrations:
50 Ohm Circuit
Parameter/Equipment | Range | Best Uncertainty (±) | Comments |
---|---|---|---|
Amplitude | (0.250 to 4) kV | 4.23% | IEC 61000-4-4 |
Rise Time | 5 μs ± 30 % | 4.40% | IEC 61000-4-4 |
Pulse Width | 50 μs ± 30 % | 4.40% | IEC 61000-4-4 |
Burst Period | 15ms | 3.50% | IEC 61000-4-4 |
1k Ohm Circuit
Parameter/Equipment | Range | Best Uncertainty (±) | Comments |
---|---|---|---|
Amplitude | (0.250 to 4) kV | 4.29% | IEC 61000-4-4 |
Rise Time | 5 μs ± 30 % | 4.35% | IEC 61000-4-4 |
Pulse Width | 50 μs ± 30 % | 4.40% | IEC 61000-4-4 |
Bust Period | 15ms | 3.58% | IEC 61000-4-4 |